Surface current density mapping for identification of gastric slow wave propagation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19403355.
- Also identified by DOI 10.1109/TBME.2009.2021576 and PMC identifier 2722927.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The magnetogastrogram (MGG) records clinically relevant parameters of the electrical slow wave of the stomach noninvasively. Besides slow wave frequency, gastric slow wave propagation velocity is a potentially useful clinical indicator of the state of health of gastric tissue, but it is a difficult parameter to determine from noninvasive bioelectric or biomagnetic measurements. We present a method for computing the surface current density from multichannel MGG recordings that allows computation of the propagation velocity of the gastric slow wave. A moving dipole source model with hypothetical as well as realistic biomagnetometer parameters demonstrates that while a relatively sparse array of magnetometer sensors is sufficient to compute a single average propagation velocity, more detailed information about spatial variations in propagation velocity requires higher density magnetometer arrays. Finally, the method is validated with simultaneous MGG and serosal electromyography measurements in a porcine subject.
Medical subject headings
- Electrophysiology
- Gastrointestinal Motility
- Models, Biological
- Stomach