Primary total hip arthroplasty with a porous-coated acetabular component. A concise follow-up, at a minimum of twenty years, of previous reports.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19411461.
- Also identified by DOI 10.2106/JBJS.H.00168.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We previously reported the seven and fifteen-year results of the use of a porous-coated acetabular metal shell inserted without cement in a consecutive series of 204 primary total hip arthroplasties. In the present study, we evaluated the longer-term outcomes of these arthroplasties at a minimum follow-up time of twenty years. One hundred and fourteen (92%) of the 124 hips available for study had retained the original acetabular metal shell. A total of five acetabular components had been revised for aseptic loosening or had radiographic evidence of definite loosening. Fourteen hips with well-fixed acetabular shells required a change of the modular acetabular liner because of excessive wear and/or for the treatment of osteolysis, and liner changes have been recommended for another eight hips. The twenty-year rate of survival of the metal shell, with failure defined as revision because of loosening or radiographic evidence of loosening, was 96% (95% confidence interval, 94% to 98%). Cementless acetabular reconstruction continues to provide durable fixation at twenty years postoperatively. Wear-related complications continue to be the major mode of failure.
Medical subject headings
- Arthroplasty, Replacement, Hip
Anatomy
- hip