Modular ulnar head decoupling force: case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19446967.
- Also identified by DOI 10.1016/j.jhsa.2009.02.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cobalt-chrome modular distal ulnar head replacement arthroplasty is a surgical option to restore stability to the distal radioulnar joint rendered unstable by hemi-resection arthroplasty or a total resection arthroplasty. However, the revision of dislocated modular cobalt-chrome ulnar head implants may pose an important intraoperative challenge. The Morse-taper disassembly force of modular ulnar head implants is not available in the current published literature. We present a case in which tremendous difficulty was encountered while revising a dislocated modular cobalt-chrome distal ulnar head implant. The mean Morse-taper disassembly force of the retrieved modular cobalt-chrome implant was 2958 N +/- 1272. At nearly 4.5 times the average body weight, the modular ulnar head Morse-taper disassembly strength presented a formidable force to overcome intraoperatively.
Medical subject headings
- Arthroplasty, Replacement
- Device Removal
- Joint Prosthesis
- Ulna
- Wrist Joint
Anatomy
- ulna
- wrist
- radius/ulna