Vertical arrays of nanofluidic channels fabricated without nanolithography.

Sordan, Roman; Miranda, Alessio; Traversi, Floriano; Colombo, Davide; Chrastina, Daniel; Isella, Giovanni; Masserini, Massimo; Miglio, Leo et al. · Lab Chip · 2009

basic_science · Level V

Where this comes from

Abstract

Vertical arrays of sealed nanofluidic channels, in which both cross-sectional dimensions are controllable down to 10 nm, were fabricated by selective side etching of a SiGe heterostructure comprised of layers of alternating Ge fractions. Capillary filling of these nanochannel arrays with fluorescent dye solutions was investigated using a confocal microscope. The feasibility of using nanochannels for size-based separation of biomolecules was demonstrated by imaging aggregates of tagged amyloid-beta peptide. The ability to integrate a large number of nanochannels shows promise for high throughput applications involving lab-on-a-chip systems.

Medical subject headings