Identification of a shine-through artifact in the trachea with (124)I PET/CT.

Abdul-Fatah, Sakar B; Zamburlini, Mariangela; Halders, Servé G E A; Brans, Boudewijn; Teule, Gerrit J J; Kemerink, Gerrit J · J Nucl Med · 2009

basic_science · Level V

Where this comes from

Abstract

The presence of (124)I in tissue near the trachea can cause a phenomenon that might be called shine-through. The effect is due to high-energy positrons that cross the air-filled trachea and annihilate at the opposite tracheal wall, incorrectly suggesting uptake at that location. We investigated shine-through using (124)I, (68)Ga, and (18)F PET/CT scans of a neck phantom. Additionally, we evaluated (124)I studies of 29 patients with differentiated thyroid cancer who underwent imaging for postsurgical staging. In the phantom studies with a 0.1-mL (124)I source, the relative intensity of shine-through decreased from 7% to nearly zero when the thickness of the positron-stopping layer was increased from 0.3 to 3.85 mm. In patients, shine-through was observed in 5 of the 29 studies, with intensities between 0.7% and 14%. Shine-through appears rather common in differentiated thyroid cancer. Recognition is important for identification of real lesions, calculation of uptake, and dosimetry.