Identification of a shine-through artifact in the trachea with (124)I PET/CT.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19470851.
- Also identified by DOI 10.2967/jnumed.108.060442.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The presence of (124)I in tissue near the trachea can cause a phenomenon that might be called shine-through. The effect is due to high-energy positrons that cross the air-filled trachea and annihilate at the opposite tracheal wall, incorrectly suggesting uptake at that location. We investigated shine-through using (124)I, (68)Ga, and (18)F PET/CT scans of a neck phantom. Additionally, we evaluated (124)I studies of 29 patients with differentiated thyroid cancer who underwent imaging for postsurgical staging. In the phantom studies with a 0.1-mL (124)I source, the relative intensity of shine-through decreased from 7% to nearly zero when the thickness of the positron-stopping layer was increased from 0.3 to 3.85 mm. In patients, shine-through was observed in 5 of the 29 studies, with intensities between 0.7% and 14%. Shine-through appears rather common in differentiated thyroid cancer. Recognition is important for identification of real lesions, calculation of uptake, and dosimetry.
Medical subject headings
- Iodine Radioisotopes
- Positron-Emission Tomography
- Tomography, X-Ray Computed
- Trachea