Anterior cruciate ligament revision of a relatively new implant system.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19472962.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cayenne AperFix (Cayenne Medical, Inc, Scottsdale, Arizona) is a relatively new implant used in anterior cruciate ligament (ACL) reconstruction. To date, there have been more than 4500 primary reconstructions performed using the device, and removal of the implant during revision surgery has not been discussed in any published literature. The focal point of the procedure involves removing the center screw within the femoral component, followed by using a core reamer to extract the entire device, leaving behind a femoral tunnel with no defects that can accept a new graft. We present the case of failed primary ACL reconstruction in a 23-year-old man due to a single traumatic event. We discuss the technical challenges associated with this implant as well as the technique used for removal.
Medical subject headings
- Anterior Cruciate Ligament
- Anterior Cruciate Ligament Injuries
- Prostheses and Implants
- Plastic Surgery Procedures