Radiation dose exposure of computed tomography coronary angiography: comparison of dual-source, 16-slice and 64-slice CT.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19482842.
- Also identified by DOI 10.1136/hrt.2008.161018.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Dual-source CT (DSCT) promises a significant reduction of radiation dose exposure for coronary CT angiography (CTA). Large studies on radiation dose estimates are rare. To compare radiation dose estimates of DSCT with 16- and 64-slice multidetector CT (MDCT) for non-invasive coronary angiography. Retrospective data analysis was performed on 292 patients: 56 patients were examined with 16-slice MDCT, 38 patients with 64-slice MDCT and 202 patients using DSCT. The effective dose (ED) estimates were calculated for all patients from the dose-length product and the conversion factor k (0.017 mSv/mGy/cm), as recommended by current guidelines. The mean (SD) ED for patients examined by 16-slice MDCT was 9.8 (1.8) mSv, for 64-slice MDCT 8.6 (2.8) mSv and for DSCT 11.4 (7.2) mSv. With a protocol of 100 kV tube voltage and 110 ms ECG pulsing window the mean (SD) ED was 3.8 (1.7) mSv for DSCT scanning. When DSCT with a tube voltage of 100 kV was used, a significant inverse correlation between heart rate and radiation dose exposure was found. When standard protocols for coronary CTA with 16-, 64-slice MDCT and DSCT scanners are used, the radiation dose is still high. However, using optimised and individually adjusted protocols low estimated radiation doses can be achieved.
Medical subject headings
- Coronary Angiography
- Coronary Artery Disease
- Tomography, X-Ray Computed