Cementless acetabular socket revisions using Metasul metal-on-metal bearings.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19493650.
- Also identified by DOI 10.1016/j.arth.2009.04.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thirty-four hips were treated with cementless acetabular socket revisions using a metal-on-metal bearing. The causes of revision were aseptic loosening in 33 hips and septic loosening in 1 hip. Revisions were performed for acetabular sockets in 28 hips and for acetabular sockets and femoral stems in 6 hips. Mean follow-up duration was 6.2 years (range, 4.0-9.1 years), and mean Harris Hip Scores improved from 56 to 92 points. No hip required further revision for aseptic loosening. Focal femoral osteolysis newly developed in zone I in 1 hip, which was treated by curettage and bone grafting. The authors suggest that second-generation metal-on-metal bearings in cementless acetabular socket revisions can achieve good medium-term clinical and radiographic results.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Biocompatible Materials
- Hip Prosthesis
- Joint Diseases
- Metals
Anatomy
- hip
- pelvis