Mechanical properties, oxidation, and clinical performance of retrieved highly cross-linked Crossfire liners after intermediate-term implantation.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19520545.
- Also identified by DOI 10.1016/j.arth.2009.04.022 and PMC identifier 2876196.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Sixty Crossfire (Stryker Orthopaedics, Mahwah, NJ) liners were consecutively revised after an average of 2.9 years (range, 0.01-8.0 years) for reasons unrelated to wear or mechanical performance of the polyethylene. Femoral head penetration was measured directly from 42 retrievals implanted for more than 1 year. Penetration rate results (0.04 mm/y, on average; range, 0.00-0.13 mm/y) confirmed decreasing wear rates with longer in vivo times. Overall, we observed oxidation levels at the bearing surface of the 60 liners (0.5, on average; range, 0.1-1.7) comparable to those of nonimplanted liners (0.5, on average; range, 0.3-1.1) and preservation of mechanical properties. We also measured elevated oxidation of the rim (3.4, on average; range, 0.2-8.8) that was correlated with implantation time. Rim surface damage, however, was observed in only 3 (5%) of 60 cases. Retrieval analysis of the 3 rim-damaged liners did not reveal an association between surface damage and the reasons for revision.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Equipment Failure Analysis
- Prosthesis Failure
Anatomy
- hip