Sterilization of medical devices by ethylene oxide, determination of the dissipation of residues, and use of Green Fluorescent Protein as an indicator of process control.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19530160.
- Also identified by DOI 10.1002/jbm.b.31438.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ethylene oxide (EO) is used to sterilize Oxygenator and Tubing applied to heart surgery. Residual levels of EO and its derivatives, ethylene chlorohydrin (ECH) and ethylene glycol (EG), may be hazardous to the patients. Therefore, it must be removed by the aeration process. This study aimed to estimate the minimum aeration time for these devices to attain safe limits for use (avoiding excessive aeration time) and to evaluate the Green Fluorescent Protein (GFP) as a biosensor capable of best indicating the distribution and penetration of EO gas throughout the sterilization chamber. Sterilization cycles of 2, 4, and 8 h were monitored by Bacillus atrophaeus ATCC 9372 as a biological indicator (BI) and by the GFP. Residual levels of EO, ECH, and EG were determined by gas chromatography (GC), and the residual dissipation was studied. Safe limits were reached right after the sterilization process for Oxygenator and after 204 h of aeration for Tubing. In the 2 h cycle, the GFP concentration decreased from 4.8 (+/-3.2)% to 7.5 (+/-2.5)%. For the 4 h cycle, the GFP concentration decreased from 17.4 (+/-3.0)% to 21.5 (+/-6.8)%, and in the 8 h cycle, it decreased from 22.5 (+/-3.2)% to 23.9 (+/-3.9)%. This finding showed the potentiality for GFP applications as an EO biosensor.
Medical subject headings
- Disinfectants
- Drug Residues
- Ethylene Oxide
- Green Fluorescent Proteins
- Sterilization