Comparison of corneal powers obtained from 4 different devices.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19541287.
- Also identified by DOI 10.1016/j.ajo.2009.04.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the repeatability and comparability of anterior corneal power values obtained from the Galilei Dual Scheimpflug Analyzer (Ziemer, Port, Switzerland), Humphrey Atlas corneal topographer (Carl Zeiss, Jena, Germany), IOLMaster (Carl Zeiss), and a manual keratometer (Bausch & Lomb Inc, Rochester, New York, USA). Prospective, comparative study. Prospectively, 20 subjects were enrolled. Three sets of corneal power measurements were obtained by a single observer using the Galilei, Atlas topographer, IOLMaster, and manual keratometer. Repeatability of the 3 measurements from each device was evaluated by means of coefficient of variation, standard deviation (SD), and intraclass correlation coefficient. An analysis of variance was used to compare the differences in corneal powers among devices. The Bland and Altman method also was performed to assess agreement in measurements between devices. Vector analysis was used to compare the astigmatism values obtained from different devices. For each device, the coefficient of variation of repeated measurements was lower than 0.22%. The SD of 3 repeated measurements ranged from 0.042 to 0.096 diopters (D). The intraclass correlation coefficients were higher than 0.99 in all devices. Mean central corneal powers were 43.80 D, 43.88 D, 43.92 D, and 43.76 D for the Galilei, Atlas, IOLMaster, and manual keratometer, respectively. SDs of the differences between devices ranged from 0.07 D for Galilei and IOLMaster to 0.14 D for Galilei and Atlas. For astigmatism, the mean astigmatism values for the Galilei, Atlas, IOLMaster, and manual keratometer were 0.54 D at 84 degrees, 0.51 D at 88 degrees, 0.62 D at 88 degrees, and 0.52 D at 87 degrees, respectively. The corneal power measurements from these 4 devices were highly reproducible, comparable, and correlated.
Medical subject headings
- Cornea
- Diagnostic Techniques, Ophthalmological
- Refraction, Ocular