Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19577896.
- Also identified by DOI 10.1016/j.arth.2009.05.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Given that the manufacture of highly cross-linked polyethylene (HXLPE) is not standardized, the behavior of these materials may vary. Our study compares minimum 5-year steady state femoral head penetration rates using the Martell method, in 2 HXPLEs produced by different manufacturers. Patients received a primary hip arthroplasty using an uncemented acetabular component with an HXLPE liner and a 28-mm femoral head. Forty-seven patients in group A received an HXLPE liner (Reflection XLPE, Smith and Nephew Inc, Memphis, Tenn), and 36 patients in group B received a different HXLPE liner (Longevity, Zimmer Inc, Warsaw, Ind). Average follow-up was 6.42 years in group A and 7.64 years in group B. The steady state head penetration rates were not significantly (P > .05) different between the HXPLE groups over the midterm with 0.026 mm/y and 0.025 mm/y in groups A and B, respectively.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur Head
- Hip Prosthesis
- Osteoarthritis, Hip
- Polyethylenes
Anatomy
- femur
- hip