A comparison of stroke volume variation measured by Vigileo/FloTrac system and aortic Doppler echocardiography.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19608819.
- Also identified by DOI 10.1213/ane.0b013e3181ac6dac.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The goal of this study was to compare stroke volume variation (SVV) assessed from a peripheral artery with the Vigileo/FloTrac system (SVV-FloTrac) with SVV derived close to the heart by aortic Doppler (SVV-Doppler). Thirty patients undergoing liver transplantation underwent simultaneous SVV-FloTrac and SVV-Doppler measurements before and after intravascular volume expansion. SVV-FloTrac and SVV-Doppler comparison before intravascular volume expansion showed a mean bias of 0.7%, and 95% limits of agreement of -4.2% to 5.5%. The areas under the receiver operating characteristic curves generated to discriminate responders and nonresponders to intravascular volume expansion were not different for SVV-FloTrac and SVV-Doppler. SVV-FloTrac and SVV-Doppler measurements show acceptable bias and limits of agreement, and similar performance in terms of fluid responsiveness in patients undergoing liver transplantation.
Medical subject headings
- Arteries
- Echocardiography, Doppler
- Stroke Volume