Easy retrieval of polyaxial tulip-head pedicle screws by "U" rod technique.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19618219.
- Also identified by DOI 10.1007/s00586-009-1096-4 and PMC identifier 2899743.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The number of fusion surgeries increase each year which also increase the need for implant removal. In some cases, it can be extremely hard to remove a pedicle screw especially when there is a mismatch of the screw and the screwdriver. Also the screwdrivers can be contaminated during the operation, and this will cause a delay till the instruments are re-sterilized. There is a need for the removal of screws without special instruments. We describe a method for removing tulip-head polyaxial pedicle screws without special instruments. The screws are removed using an Allen key, a rod bender and a "U" shaped rod. We successfully removed 76 screws in 11 recent cases without any complications. The "U" rod technique is a simple and useful technique for the removal of tulip-head polyaxial screws.
Medical subject headings
- Bone Screws
- Spinal Diseases
- Spinal Fusion
- Spine
- Surgical Instruments