Bacterial contamination of PCA and epidural infusion devices.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19624630.
- Also identified by DOI 10.1111/j.1365-2044.2009.05914.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We prospectively audited the bacterial contamination of re-usable analgesia infusion pumps. In a one-month period, 112 samples from the handset and keypads of our analgesia infusion pumps were cultured for bacterial contamination. Forty-five percent of handset swabs and 46% of keypad swabs grew bacteria; the commonest organism being coagulase-negative staphylococcus. An additional cleaning step using 70% isopropyl alcohol wipes was introduced and the contamination rate was re-audited in 100 samples. The contamination rate was reduced to 6% of handset swabs and 4% of keypad swabs. A high initial rate of bacterial contamination of re-usable analgesia infusion pumps was significantly reduced by the implementation of a simple, additional cleaning procedure.
Medical subject headings
- Analgesia, Epidural
- Analgesia, Patient-Controlled
- Bacteria
- Equipment Contamination
- Infusion Pumps