Regional anaesthesia elastomeric pump performance after a single use and subsequent refill: a laboratory study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19624633.
- Also identified by DOI 10.1111/j.1365-2044.2009.05941.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ambulatory local anaesthetic delivery systems are often limited by a short effective duration of infusion. Prolonging nerve blockade by substituting a new pump as recommended by the manufacturers, represents a substantial consumable item cost ($US 300-500). We therefore evaluated the flow delivery performance of 31 single model elastomeric devices (all with a 2 ml.h(-1) background and 5 ml every hour bolus capability) that had been filled, used in clinical practice and then refilled in the laboratory. For the second infusion, there was a pattern of over-infusion (< 10 ml.h(-1)) in the first hour; however, all pumps depreciated to < 150% of predicted by the second hour. The subsequent performance of all pumps was not only within safe limits, but also predominantly within the range (background infusion +/- 15%, bolus +10/-20%) specified by the manufacturer for primary infusion. We conclude that this elastomeric regional anaesthesia pump design performs satisfactorily after having been refilled following a single previous use.
Medical subject headings
- Anesthesia, Conduction
- Infusion Pumps