A clinical and radiostereometric study of the cemented PFC-sigma prosthesis: a 5-year study of 29 cases with a fixed bearing.

von Schewelov, Thord; Besjakov, Jack; Sanzén, Lennart; Carlsson, Ake · J Knee Surg · 2009

case_series · Level IV

Where this comes from

Abstract

The press-fit condylar (PFC) cruciate-retaining total knee prosthesis is well documented in the literature. In 1997, a modification of the femoral component was introduced, and the prosthesis was renamed PFC-Sigma. The alteration may influence the migratory and rotational pattern of the tibial component and thus long-time survival rate. In this radiostereometric analysis, we found that the micromotion of the PFC-Sigma prosthesis differs slightly from the original PFC design, an advantage of the PFC-Sigma prosthesis. The median maximum total point motion at 5 years was 0.64 mm for the PFC-Sigma design and 0.79 mm for the previous version of PFC (P = .9). However, the PFC-Sigma rotated less around the transverse (x) axis than did the PFC (medians, 0.22 mm and 0.48 mm, respectively; P = .04). From the radiostereometric, radiographic, and clinical data, we conclude that the PFC-Sigma knee prosthesis can be used with confidence.