Transcricothyroid electromyographic monitoring of the recurrent laryngeal nerve.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19650135.
- Also identified by DOI 10.1002/lary.20228.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the usefulness of intraoperative nerve monitoring using an electrode placed in the midline through the cricothyroid membrane. Retrospective records review. Patients of the otolaryngology department of our tertiary care academic medical center were identified if they had undergone either total thyroidectomy or total thyroid lobectomy accompanied by bilateral electromyographic (EMG) monitoring of the recurrent laryngeal nerve (RLN) between January 2007 and October 2007. Forty-three patients were identified who had a total thyroidectomy or a total thyroid lobectomy. Nineteen of the 43 had intraoperative EMG monitoring of the RLN with placement of a single EMG electrode through the cricothyroid membrane; 11 had a thyroidectomy, and eight had a thyroid lobectomy. Overall, 30 at-risk nerves were stimulated with an EMG probe; 27 responded adequately and three (paralyzed before surgery) were unresponsive. Central placement of an EMG electrode through the cricothyroid membrane into the thyroarytenoid musculature is a safe and reliable technique for bilateral monitoring of the RLN that facilitates evaluation of postoperative function.
Medical subject headings
- Electromyography
- Intraoperative Complications
- Monitoring, Intraoperative
- Recurrent Laryngeal Nerve Injuries
- Thyroidectomy