The effect of ECG interference on pattern-recognition-based myoelectric control for targeted muscle reinnervated patients.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19692302.
- Also identified by DOI 10.1109/TBME.2008.2010392.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Targeted muscle reinnervation has been introduced as an effective neural machine interface. In the case of a shoulder disarticulation patient, an effective site for a nerve transfer involves the pectoralis muscles, as these perform little useful function with a missing limb. Consequently, the myoelectric signals measured from the reinnervated muscles may be corrupted by a large amount of ECG interference. This paper investigates the effect of ECG upon the accuracy of a pattern-classification-based scheme for myoelectric control of powered upper limb prostheses. The results suggest that ECG interference, at levels typically encountered in a clinical measurement, has little effect upon classification accuracy, but can affect the estimate of myoelectric activity used to convey the velocity of motion (commonly referred to as proportional control). High-pass filtering at approximately 100 Hz appears to effectively mitigate the effect of ECG interference.
Medical subject headings
- Artificial Limbs
- Electrocardiography
- Electromyography
- Pattern Recognition, Automated