Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 19698909.
- Also identified by DOI 10.1016/j.arth.2009.05.027.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This prospective, randomized protocol evaluated femoral head penetration after total hip arthroplasty in a young population. Forty-five patients randomly received either a cross-linked or conventional ultrahigh-molecular-weight polyethylene (UHMWPE) liner in a noncemented hemispheric cup (Trilogy, Zimmer, Warsaw, Ind) with a 28-mm femoral head. Radiostereometric analysis film pairs, Harris hip, UCLA, SF-12, and Western Ontario and McMaster Universities scores were obtained through 2 years. Median femoral head penetration was less among cross-linked compared to conventional liners as follows: 0.06 mm (0.04-0.08 mm) vs 0.08 mm (0.02-0.19 mm) at 6 months, 0.07 mm (-0.14 to 0.16 mm) vs 0.11 mm (0.01-0.27 mm) at 1 year, and 0.065 mm (-0.04 to 0.193 mm) vs 0.169 mm (0.09-0.22 mm) at 2 years. Clinical outcomes were similar between the groups. Highly cross-linked UHMWPE demonstrated 55% less femoral head penetration compared to conventional polyethylene at 2 years. Despite improvements in the manufacturing process and sterilization of conventional UHMWPE, the femoral head penetration rate is unchanged from historical standards.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur Head
- Polyethylene
- Prosthesis Failure
Anatomy
- femur
- hip