Assessment of trabecular bone structure using MDCT: comparison of 64- and 320-slice CT using HR-pQCT as the reference standard.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19711081.
- Also identified by DOI 10.1007/s00330-009-1571-7 and PMC identifier 2814042.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of our study was to perform trabecular bone structure analysis with images from 64- and 320-slice multidetector computed tomography (MDCT) and to compare these with high-resolution peripheral computed tomography (HR-pQCT). Twenty human cadaver distal forearm specimens were imaged on a 64- and 320-slice MDCT system at 120 kVp, 200 mA and 135 kVp, 400 mA (in-plane pixel size 234 microm; slice thickness 500 microm). HR-pQCT imaging was performed at an isotropic voxel size of 41 microm. Bone volume fraction (BV/TV), trabecular number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp) were computed. MDCT-derived BV/TV and Tb.Sp were highly correlated (r = 0.92-0.96, p < 0.0001) with the corresponding HR-pQCT parameters. Tb.Th was the only structure measure that did not yield any significant correlation. The 64- and 320-slice MDCT systems both perform equally well in depicting trabecular bone architecture. However, because of constrained resolutions accurate derivation of trabecular bone measures is limited to only a subset of microarchitectural parameters.
Medical subject headings
- Bone and Bones
- Tomography, X-Ray Computed