Influence of technique with distally fixed modular stems in revision total hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19729272.
- Also identified by DOI 10.1016/j.arth.2009.07.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Distally fixed modular implants have seen a recent increase in use, to manage proximal femoral bone loss often encountered during revision total hip arthroplasty (THA). Forty-three distally fixed modular stems implanted at our institution between 1999 and 2006 were clinically and radiographically reviewed. These patients had either a minimum 2-year follow-up (average, 2.4 years; range, 2-5.6 years) or failure (ie, explant or rerevision required). Eleven stems subsided, and 4 were rerevised (n = 4), for a rate of 9.3%. All revised stems were radiographically undersized, emphasizing the importance of the technique. Although being a valuable option in revision THA, these stems are not free of complications. The high rate of subsidence encountered in our early experience shows that there is a learning curve. This complication is preventable by avoiding undersizing.