Randomised comparison of Pentax AirwayScope and Glidescope for tracheal intubation in patients with normal airway anatomy.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 19735405.
- Also identified by DOI 10.1111/j.1365-2044.2009.06032.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We compared intubating characteristics of the Pentax AirwayScope and Glidescope in a randomised controlled trial involving 140 patients. We found significantly shorter intubation times with mean (SD) 21.3 (12.3) vs 30.2 (13.2) s, lower intubating difficulty scores 4.4 (10.4) vs 12.8 (16.3) p < 0.001, and better grade 1 laryngeal views with the Pentax AirwayScope (95.7 vs 81.4%, p = 0.015). Significantly more optimisation manoeuvres were needed to successfully intubate with the Glidescope, with significant inability to align the tip of the tracheal tube with the glottic opening with the Glidescope in 9 (13%) vs no patients, p = 0.013. There was more mucosal bleeding (4 (5.7%) vs 1 (1.4%), p = 0.366), more lip bleeding (3 (4.3%) vs 1 (1.4%), p = 0.620) and significantly more postoperative sorethroat, 13 (18.6%) vs 0, p < 0.001 associated with the Glidescope.
Medical subject headings
- Intubation, Intratracheal
- Laryngoscopes