Efficacy of a cooled bipolar epicardial radiofrequency ablation probe for creating transmural myocardial lesions.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19748103.
- Also identified by DOI 10.1016/j.jtcvs.2009.06.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Creation of transmural myocardial lesions with epicardial surgical devices to treat atrial fibrillation is difficult. A new cooled bipolar radiofrequency ablation probe was used to create transmural myocardial lesions under controlled conditions. The Coolrail (AtriCure, Inc, West Chester, Ohio) is a handheld probe with 2 parallel 30-mm long radiofrequency conductors. Conductors are cooled by water irrigation. Lesions were delivered to epicardial surface of isolated bovine myocardium sliced 3- to 8-mm thick, with blood flow beneath tissue at 0 or 0.4 m/s. Contact pressure between probe and tissue was either 450 g or 900 g. Tissue temperatures were measured. Tissue was sectioned every 5 mm along lesion long axis to determine lesion dimensions. For 80 experiments with 450-g contact pressure, epicardial lesion length was 31.3 mm (interquartile range, 30.1-32.8 mm); endocardial lesion length was 14.1 mm (interquartile range, 0.0-22.6 mm). Average lesion depth was 4.2 +/- 0.74 mm. Temperature at probe interface was 81 degrees C +/- 21 degrees C; that at blood pool interface was 53 degrees C +/- 12 degrees C. Lesions were always transmural when tissue thickness was 4.0 mm or less. Endocardial blood flow did not influence lesion depth. With 900-g contact pressure, increased depth was always transmural at 4.8-mm tissue thickness or less. This irrigated bipolar radiofrequency probe consistently produced transmural lesions in tissue 4 mm or thinner under controlled conditions in vitro. Lesion depth was increased by greater pressure on probe and not affected by blood flow. Endocardial lesions were smaller than epicardial dimensions.
Medical subject headings
- Catheter Ablation