A controllably anisotropic conductivity or diffusion phantom constructed from isotropic layers.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19760146.
- Also identified by DOI 10.1007/s10439-009-9799-6 and PMC identifier 2809807.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Phantoms with controllable and well-defined anisotropy are needed to test methods for imaging electrical anisotropy. We developed and tested a phantom that had properties similar to a homogeneous anisotropic conductive medium. The phantom was constructed with alternate slices of isotropic gel having different conductivities. The degree of anisotropy in the phantom could be varied easily by changing the relative conductivity of the two gels. We tested the stability of several phantoms and found their properties were maintained for approximately 8 h following construction. The phantom has application to electrical impedance tomography, magnetic resonance electrical impedance tomography, EEG and ECG source imaging and diffusion tensor imaging.
Medical subject headings
- Diagnostic Imaging
- Gels
- Phantoms, Imaging
- Plethysmography, Impedance