Correlation between screening for obstructive sleep apnea using a portable device versus polysomnography testing in a commercial driving population.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19786903.
- Also identified by DOI 10.1097/JOM.0b013e3181b68d52.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the accuracy of portable monitoring for obstructive sleep apnea (OSA) with polysomnography (PSG) in commercial drivers. A consecutive case series of drivers were screened for OSA using recent validated consensus conference criteria at a single occupational medicine clinic; those screening positive were tested with a portable OSA screening device while awaiting PSG. Three hundred forty-six drivers were screened and 34 completed PSG and portable monitoring. Comparing the portable device to PSG at a definition of OSA as an apnea hypopnea index (AHI) >or=15, positive predictive value is 0.64 with a 0.87 negative predictive value. Positive likelihood ratio is 4.20. Linear regression model found statistically significant relationship (P = 0.0004) between the portable device and PSG AHI values. Although not perfect, portable monitoring may be useful in further stratifying risk in drivers who have a high probability of having OSA.
Medical subject headings
- Automobile Driving
- Monitoring, Ambulatory
- Occupations
- Polysomnography
- Sleep Apnea, Obstructive