Comparison of different sizes of bioabsorbable interference screws for anterior cruciate ligament reconstruction using bioabsorbable bead augmentation in a porcine model.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19801288.
- Also identified by DOI 10.1016/j.arthro.2009.05.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to compare the initial fixation strength of tendon grafts between different sizes of bioabsorbable interference screws (BioScrew; Linvatec, Largo, FL) with bioabsorbable bead (EndoPearl; Linvatec) augmentation through biomechanical analysis of a porcine femoral bone model. Forty pairs of porcine femurs and porcine flexor digitorum profundus tendons were divided into control and study groups. In the control group 8 x 30-mm BioScrews alone (n = 10) were inserted, whereas different sizes of BioScrews, measuring 7 x 30 mm (n = 10), 8 x 30 mm (n = 10), and 9 x 30 mm (n = 10), with 8-mm EndoPearl augmentation were inserted individually for fixation of tendon grafts in the study groups. All specimens were cyclically loaded with axial forces between 50 and 250 N at 1 Hz for 3,000 cycles and then incrementally loaded to failure at a rate of 150 mm/min. BioScrews with EndoPearl augmentation had a significantly higher failure load than BioScrews alone (8-mm BioScrew alone v 8-mm BioScrew and EndoPearl, P < .05). There were no significant differences in the ultimate failure load (8 mm v 7 mm and 9 mm, P = .201 and P = .871, respectively), stiffness (8 mm v 7 mm and 9 mm, P = .789 and P = .823, respectively), displacement (8 mm v 7 mm and 9 mm, P = .695 and P = .781, respectively), and bone mineral density (P = .728 for all comparisons) except insertion torque (8 mm v 7 mm and 9 mm, P = .045 and P = .518, respectively) between study groups. Less tendon laceration by the screw thread was noted in the group in which smaller-sized BioScrews were used. When EndoPearl augmentation was used, smaller-sized BioScrews (BioScrew size 1 mm smaller than bone tunnel) offered equivalent graft fixation strength to BioScrews of similar or larger sizes. Smaller-sized BioScrews can be chosen if EndoPearl augmentation has been used, and EndoPearl augmentation may reduce the risk of tendon rupture while BioScrews are inserted.
Medical subject headings
- Absorbable Implants
- Anterior Cruciate Ligament
- Bone Screws
- Implants, Experimental
- Plastic Surgery Procedures
Anatomy
- femur