Sequential activation of ground pads reduces skin heating during radiofrequency tumor ablation: in vivo porcine results.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19822466.
- Also identified by DOI 10.1109/TBME.2009.2033385 and PMC identifier 3108072.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Skin burns below ground pads during monopolar RF ablation are increasingly prevalent, thereby hindering the development of higher power RF generators capable of creating larger tumor ablation zones in combination with multiple or new applicators. Our goal was to evaluate reduction in skin temperatures via additional ground pads in an in vivo porcine model. Three ground pads placed on the animal's abdomen were activated either simultaneously or sequentially, where activation timing was adjusted to equilibrate skin temperature below each pad. Thirteen RF ablations (n = 4 simultaneous at 300 W, n = 5 sequential at 300 W, and n = 4 sequential at 375 W) were performed for 12 min via two internally cooled cluster electrodes placed in the gluteus maximus of domestic swine. Temperature rise at each pad and burn degree as determined via histology were compared. Ablation zone size was determined via T2-weighted MRI. Maximum temperature rise was significantly higher with simultaneous activation than with either of the sequential activation group (21.4 degrees C versus 8.1 degrees C or 9.6 degrees C, p < 0.01). Ablation zone diameters during simultaneous (300 W) and sequential activations (300 and 375 W) were and 6.9 +/- 0.3, 5.6 +/- 0.3, and 7.5 +/- 0.6 cm, respectively. Sequential activation of multiple ground pads results in significantly lower skin temperatures and less severe burns, as measured by histological examination.
Medical subject headings
- Burns
- Catheter Ablation
- Skin
- Skin Temperature