Ultrasound-based navigation and 3D CT compared in acetabular cup position.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19835299.
- Also identified by DOI 10.3928/01477447-20090915-50.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Intraoperative landmarks are used in image-free navigation systems. The ultrasound-based navigation systems try to overcome the problems of positional deviation associated with soft tissue. Our study analyzed the accuracy of ultrasound-based navigation of cup positioning compared with postoperative 3-dimensional (3D) computed tomography scans of cup positioning. Twenty-five ultrasound-navigated total hip arthroplasties (THAs) were analyzed. The difference between the intraoperative cup orientation (navigation) and the postoperative cup position (CT) was evaluated. The average difference between intraoperative navigation and postoperative CT measurements was 2.8 degrees (SD+/-1.8 degrees ) for abduction and 2.2 degrees (SD+/-1.6 degrees ) for anteversion. Therefore, we recommend ultrasound-based navigation as an exact tool for cup positioning in THA.