Accurate automatic exposure controller for mammography: design and performance.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1987599.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An automatic exposure controller has been designed that controls the optical film density for film, screen, and radiographic techniques typically used in mammography to within 0.05 over a range of 1.3-6.7-cm thickness of Lucite. This degree of accuracy is better than that reported for presently available controllers. The detector system consists of four cadmium telluride detectors and involves the use of a control algorithm to read the detectors and turn off the mammography unit at the correct time. This algorithm is implemented by a microprocessor, which also provides the means for a convenient calibration.