Minimum five-year follow-up wear measurement of longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 19879725.
- Also identified by DOI 10.1016/j.arth.2009.09.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated the efficacy of combining highly cross-linked polyethylene with ceramic heads on further reduction in polyethylene wear compared with the combination with cobalt-chromium heads via PolyWare computer-assisted method. A prospective cohort study was performed on 102 cementless total hip arthroplasties using Longevity (Zimmer, Warsaw, Ind) highly cross-linked polyethylene liners. Either 26-mm zirconia heads or 26-mm cobalt-chromium heads were randomly used in 51 hips each. At a mean follow-up of 6.7 years, no significant differences were identified between the groups for total penetration rate and steady-state wear rate. Osteolysis was not observed in any hips in either group. In conclusion, no advantage was seen for the 26-mm zirconia head compared with the 26-mm cobalt-chromium head in this period.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium Alloys
- Hip Prosthesis
- Polyethylene
- Zirconium
Anatomy
- hip
- femur