Abstract
The sensitivity of technetium-99m- (99mTc) d,l-HMPAO to radiolytically induced dissociation in aqueous solutions was investigated. It was found that cobalt-60 (60Co) gamma irradiation of solutions containing 99mTc-d,l-HMPAO with only 1600 cGy reduced the lipophilic chelates' radiochemical purity (RCP) to 50%-60%. The radiolytic sensitivity of 99mTc-meso-HMPAO is significantly lower. The results indicate that radiolytically produced intermediates limit the in vitro stability of 99mTc-d,l-HMPAO.