Optimizing fresh gas flow and circuit design for the delivery of continuous positive airway pressure.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1989761.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To examine the effect of varying circuit design and the fresh gas flow rate on the circuit work imposed by a continuous positive airway pressure (CPAP) circuit. Circuit work was measured during simulated inspiration (500 mL) with a lung model at inspiratory flow rates (V) of 40, 60, and 80 L/min during the administration of 10 cm H2O CPAP through either a modified Mapleson-A or modified Mapleson-D circuit, both alone and when connected to a face mask (i.e., simulating an intubated and nonintubated patient). Fresh gas flow was varied from 10 to 250 L/min. The minimum circuit work occurred at a fresh gas flow rate approximating V; however, circuit work was consistently lower for the modified Mapleson-A circuit compared with the modified Mapleson-D circuit. As the fresh gas flow rate was increased sequentially to 250 L/min, circuit work remained close to the minimum value for the modified Mapleson-A, but increased gradually with the modified Mapleson-D, e.g., from 0.017 kg.m/L at a fresh gas flow rate and V of 80 L/min to 0.035 kg.m/L at a fresh gas flow rate of 250 L/min and a V of 80 L/min. Rotation of the fresh gas flow inlet did not change the circuit work vs. fresh gas flow rate relationship. Addition of a face mask resulted in a smaller increase in circuit work for the modified Mapleson-D with increasing fresh gas flow rate. However, unlike the modified Mapleson-A circuit alone, the addition of a mask caused circuit work to increase with increasing fresh gas flow rate. The modified Mapleson-A circuit at a fresh gas flow rate equal to V minimizes circuit work, and hence represents an optimal CPAP circuit. The increases in circuit work at fresh gas flow rates above V that were found with the modified Mapleson-D circuit are not due to inertial differences, and are likely due to turbulent gas flow.
Medical subject headings
- Positive-Pressure Respiration