A dedicated two-channel phased-array receiver coil for high-resolution MRI of the rat knee cartilage at 7 T.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19932985.
- Also identified by DOI 10.1109/TBME.2008.2006015.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In the field of small animal studies, the array coil imaging has become increasingly important. In this paper, a dedicated two-channel array coil operating at 300 MHz (7 T) for high-resolution MRI (HR-MRI) of the rat knee cartilage is presented. The average gain in signal-to-noise ratio (SNR) compared to a 15-mm multipurpose surface coil was 2.2. This SNR gain was used to improve the spatial resolution of 3-D acquisitions by decreasing the voxel size from 59 x 59 x 156 microm(3) to 51 x 51 x 94 microm(3) without time penalty. Also, a set of two array coils was used to perform a simultaneous acquisition of both knee joints of a rat, maintaining the same scanning time without SNR or spatial resolution degradation compared to the single knee joint acquisition. This two-channel array coil is a key element to perform HR-MRI and extract cartilage morphological parameters such as thickness and volume.
Medical subject headings
- Cartilage, Articular
- Hindlimb
- Image Enhancement
- Magnetic Resonance Imaging
- Magnetics
- Transducers