The proximal fixation strength of modern EVAR grafts in a short aneurysm neck. An in vitro study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19939708.
- Also identified by DOI 10.1016/j.ejvs.2009.10.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The study aims to measure the strength of the proximal fixation of endografts in short and long necks. Three types of endografts were compared: Gore Excluder, Vascutek Anaconda and Medtronic Endurant. The proximal part of the stent grafts was inserted in bovine arteries and the graft was then attached to a tensile testing machine. The force to obtain dislodgement (DF) from the aorta was recorded for each graft at proximal seal lengths of 10 and 15 mm. The median DF (interquartile range, IQR) for the Excluder, the Anaconda and the Endurant with a seal length of 15 mm was: 11.8 (10.5-12.0) N, 20.8 (18.0-30.1) N and 10.7 (10.4-11.3) N. With the shorter proximal seal of 10mm, DF was, respectively: 6.0 (4.5-6.6) N, 17.0 (11.2-36.6) N and 6.4 (6.1-12.0) N. The proximal fixation of the Anaconda is superior to the Excluder and the Endurant at short necks of 10 and 15 mm in an experimental set-up. There is a statistically significant decrease of proximal fixation for the Excluder stent graft, when decreasing the length of the proximal neck from 15 to 10 mm.
Medical subject headings
- Aortic Aneurysm
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation