Three dimensional evaluation of the aortic annulus using multislice computer tomography: are manufacturer's guidelines for sizing for percutaneous aortic valve replacement helpful?
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19995874.
- Also identified by DOI 10.1093/eurheartj/ehp534.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the effects of applying current sizing guidelines to different multislice computer tomography (MSCT) aortic annulus measurements on Corevalve (CRS) size selection. Multislice computer tomography annulus diameters [minimum: D(min); maximum: D(max); mean: D(mean) = (D(min) + D(max))/2; mean from circumference: D(circ); mean from surface area: D(CSA)] were measured in 75 patients referred for percutaneous valve replacement. Fifty patients subsequently received a CRS (26 mm: n = 22; 29 mm: n = 28). D(min) and D(max) differed substantially [mean difference (95% CI) = 6.5 mm (5.7-7.2), P < 0.001]. If D(min) were used for sizing 26% of 75 patients would be ineligible (annulus too small in 23%, too large in 3%), 48% would receive a 26 mm and 12% a 29 mm CRS. If D(max) were used, 39% would be ineligible (all annuli too large), 4% would receive a 26 mm, and 52% a 29 mm CRS. Using D(mean), D(circ), or D(CSA) most patients would receive a 29 mm CRS and 11, 16, and 9% would be ineligible. In 50 patients who received a CRS operator choice corresponded best with sizing based on D(CSA) and D(mean) (76%, 74%), but undersizing occurred in 20 and 22% of which half were ineligible (annulus too large). Eligibility varied substantially depending on the sizing criterion. In clinical practice both under- and oversizing were common. Industry guidelines should recognize the oval shape of the aortic annulus.
Medical subject headings
- Aortic Valve
- Aortic Valve Stenosis
- Tomography, X-Ray Computed