Unexpected finding of a fractured metal prosthetic femoral head in a nonmodular implant during revision total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20022458.
- Also identified by DOI 10.1016/j.arth.2009.02.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Though there are many reports of fractured femoral components after total hip arthroplasty; there are no reports of a fractured metal femoral head. This is a report of a fractured metal femoral head in a nonmodular total hip replacement discovered unexpectedly during revision total hip arthroplasty for acetabular failure. This surprise finding, which was not appreciated on preoperative x-rays, required unplanned extraction of a well-ingrown, fully porous coated cylindrical femoral stem. Though rare, fracture of the metal femoral head in a DePuy (Warsaw, Ind.) monoblock Anatomic Medullary Locking (AML) component is possible, and one cannot expect the fracture to be apparent on preoperative radiographs as the 2 pieces may not necessarily dissociate. As usual, the surgeon performing revision arthroplasty should be prepared to revise all components.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- femur
- hip