Lesion size inaccuracies in digital mammography.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 20028882.
- Also identified by DOI 10.2214/AJR.09.2927.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We show with both a clinical case and measurements using the American College of Radiology mammography phantom that some digital mammography acquisition and image display combinations lead to both marked over- and underestimation of lesion sizes on geometrically magnified images. The results of this study indicate that the accuracy of lesion size measurements in all magnification modes should be a routine quality control acceptance test with each acquisition-display system combination in digital mammography.