Imaging of soft tissues adjacent to orthopedic hardware: comparison of 3-T and 1.5-T MRI.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20028892.
- Also identified by DOI 10.2214/AJR.08.1740.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to compare metal artifact reduction techniques at 1.5-T and 3-T MRI. A titanium plate with steel screws was placed in a freshly harvested pig leg. The leg was imaged with 1.5-T and 3-T MRI. A T2-weighted turbo spin-echo sequence was used with echo-train lengths of 8, 16, 32, and 64 and a constant readout bandwidth of 31.2 kHz. The images were compared qualitatively, and the optimal echo-train length was selected. Images were acquired at the optimal echo-train length with four different readout bandwidths. Artifact was measured quantitatively, and image quality was ranked qualitatively. The qualitatively best image acquired at 1.5 T was compared with the qualitatively highest-ranked image acquired at 3 T. At both 1.5 T and 3 T, optimal images of equal quality were produced at echo-train lengths of 8 and 16. At higher readout bandwidths, there was quantitatively less artifact. The qualitatively best images were acquired at a readout bandwidth of 31.2 kHz at 1.5 T and 62.5 kHz at 3 T (Cronbach's alpha=1.00). The optimal image at 3 T was qualitatively superior to that at 1.5 T. Optimizing image acquisition parameters in this phantom model resulted in similar quantitative susceptibility artifact at 3 T and 1.5 T and better qualitative images at 3 T than at 1.5 T.
Medical subject headings
- Artifacts
- Bone Plates
- Bone Screws
- Image Enhancement
- Magnetic Resonance Imaging