Intracranial migration of a fractured intrathecal catheter from a baclofen pump system: case report and analysis of possible causes.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 20087131.
- Also identified by DOI 10.1227/01.NEU.0000363183.78323.D7.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This case report describes a new complication associated with a baclofen pump in which its fractured intrathecal catheter migrated into the patient's ventricular system. A thecal model was developed to evaluate catheter buoyancy in artificial cerebrospinal fluid (CSF). The literature was reviewed to identify possible mechanical and physiologic causes of catheter migration. A 16-year-old boy with cerebral palsy presented with cervical pain, nausea, and vomiting. He was known to have a nonfunctioning baclofen pump with a 1-piece intrathecal catheter. Imaging studies showed mild ventriculomegaly and a fractured segment of the intrathecal catheter that extended from the cervical subarachnoid space into the third and fourth ventricles. The patient had complete symptom resolution after undergoing urgent surgical removal of the catheter segment. Manufacturer analysis of the retrieved catheter revealed a crushed, jagged proximal end. In an experimental thecal sac model, catheter segments in lengths of 0.5 to 89 cm were denser than the artificial CSF and, therefore, did not float in the thecal sac. This finding negates the role of buoyancy in migration. Review of the literature advocates for caudocranial CSF flow patterns as a plausible mechanism for migration. This complication alerts surgeons to the migration risk of loose intrathecal catheter segments into the ventricular system. CSF flow patterns and mechanical processes, but not material properties of the catheter, are likely causes.
Medical subject headings
- Equipment Failure
- Foreign-Body Migration
- Infusion Pumps, Implantable
- Spinal Cord Injuries