Degradation of and tissue reaction to biodegradable poly(L-lactide) for use as internal fixation of fractures: a study in rats.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2009343.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Samples of high-molecular-weight poly(L-lactide) (PLLA) (Mv = 9.0 x 10(5), a biomaterial developed for plates and screws used in internal fixation of jaw fractures, were implanted subcutaneously in the backs of rats to study tissue reaction to PLLA and to follow the degradation process. The PLLA seemed to follow the degradation pattern typical of biodegradable polyesters. After pure hydrolysis up to about 104 wk, phagocytic activity of macrophages was found at about 143 wk. Full resorption of PLLA was not demonstrated in this study. Except for the early and final parts of the implant period, no acute or chronic inflammatory reaction was observed. No implant was rejected. It is estimated that more than 3 yr will be required for total resorption of PLLA. For bone-healing this long period is of no practical importance. There is no need for removal of PLLA after fracture healing as is the case with metal fixation devices. Thus, PLLA has potential application in internal fixation of fractures and osteotomies in the maxillofacial region and other fractures that are not too heavily loaded in the human body.
Medical subject headings
- Biocompatible Materials
- Foreign-Body Reaction
- Fracture Fixation, Internal
- Polyesters
- Prostheses and Implants
- Wound Healing