Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up.

Fukui, Kiyokazu; Kaneuji, Ayumi; Sugimori, Tanzo; Ichiseki, Toru; Kitamura, Kenji; Matsumoto, Tadami · J Arthroplasty · 2011

prospective_cohort · Level II

Where this comes from

Abstract

Highly cross-linked polyethylene (HXLPE) was developed to reduce wear of articular bearing surface in total hip arthroplasty patients. Several studies have shown reduced wear of HXLPE compared with conventional polyethylene; however, these studies had used HXLPE in combination with a Co-Cr metal head. The purpose of this study was to compare the 5-year in vivo wear of HXLPE with that of conventional PE using a zirconia femoral head. Forty-five hips with a Trilogy HXLPE (Zimmer, Warsaw, Ind) were matched and compared with a control group of 20 conventional Trilogy PE hips. The 2-dimensional linear wear rate was significantly less in the HXLPE group between 1 and 5 years postoperation (P < .001). The results show that HXLPE reduces short-term polyethylene wear against not only a Co-Cr head but also a zirconia head.