Design and optimization of coin-shaped microreactor chips for PET radiopharmaceutical synthesis.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20124050.
- Also identified by DOI 10.2967/jnumed.109.065946 and PMC identifier 3000316.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An integrated elastomeric microfluidic device, with a footprint the size of a postage stamp, has been designed and optimized for multistep radiosynthesis of PET tracers. The unique architecture of the device is centered around a 5-microL coin-shaped reactor, which yields reaction efficiency and speed from a combination of high reagent concentration, pressurized reactions, and rapid heat and mass transfer. Its novel features facilitate mixing, solvent exchange, and product collection. New mixing mechanisms assisted by vacuum, pressure, and chemical reactions are exploited. The architecture of the reported reactor is the first that has allowed batch-mode microfluidic devices to produce radiopharmaceuticals of sufficient quality and quantity to be validated by in vivo imaging. The reactor has the potential to produce multiple human doses of (18)F-FDG; the most impact, however, is expected in the synthesis of PET radiopharmaceuticals that can be made only with low yields by currently available equipment.
Medical subject headings
- Microfluidic Analytical Techniques
- Positron-Emission Tomography
- Radiopharmaceuticals