Stability of poly(3,4-ethylene dioxythiophene) materials intended for implants.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20127989.
- Also identified by DOI 10.1002/jbm.b.31597.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study presents experiments designed to study the stability of the conducting polymer poly(3,4-ethylene dioxythiophene) (PEDOT), under simulated physiological conditions using phosphate-buffered saline (PBS) and hydrogen peroxide (H(2)O(2)) (0.01 M) at 37 degrees C over a 5- to 6-week period. Voltage pulsing in PBS was used as an additional test environment. The influence of switching the counter ion used in electropolymerization from polystyrene sulphonate (PSS) to heparin was investigated. Absorbance spectroscopy and cyclic voltammetry were used to evaluate the material properties. Most of the samples in H(2)O(2) lost both electroactivity and optical absorbance within the study period, but PEDOT:PSS was found slightly more stable than PEDOT:heparin. Polymers were relatively stable in PBS throughout the study period, with around 80% of electroactivity remaining after 5 weeks, disregarding delamination, which was a significant problem especially for polymer on indium tin oxide substrates. Voltage pulsing in PBS did not increase degradation. The counter ion influenced the time course of degradation in oxidizing agents.
Medical subject headings
- Bridged Bicyclo Compounds, Heterocyclic
- Coated Materials, Biocompatible
- Materials Testing
- Polymers