Modified occlusion tests for the Bain breathing system.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 2014900.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Undetected defects in the inner tube of the Bain coaxial anaesthetic breathing system may result in a greatly increased apparatus deadspace. Several authorities have advocated tests intended to detect inner tube problems; however, the efficacy of these tests has never been validated. In this study none of the tests were able to detect all the induced defects. A modification of an existing test using the backbar pressure-relief valve and a new double occlusion test were sufficiently sensitive to detect all defects.
Medical subject headings
- Anesthesia, Inhalation
- Respiration