A platinum-chromium steel for cardiovascular stents.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20181394.
- Also identified by DOI 10.1016/j.biomaterials.2010.01.146.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The desire to reduce the strut thickness of cardiovascular stents has driven the development of a new high strength radiopaque alloy, based on additions of platinum to a chromium-rich iron based matrix. This paper reports on initial development of the alloy and the rationale for selection of the composition. Data is presented for tensile and microstructural characterization, surface oxide analysis, corrosion resistance and endothelial cell response of the alloy. The results demonstrate the solid solution strengthening effect of the platinum, with an average yield strength of 480 MPa achieved. The material surface consists of primarily chromium oxide which contributes to the high corrosion resistance observed. The cell assay result suggests that surfaces of this Pt-enhanced alloy endothelialize in a manner comparable to stainless steel.
Medical subject headings
- Alloys
- Cardiovascular System
- Steel
- Stents