Dosimetry and physical treatment planning for iodine eye plaque therapy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2022510.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The dosimetry of eye plaques loaded with iodine-125 seeds (type 6702) was performed by means of computer calculations and measurements with thermoluminescent dosimeters (TLD). Measurements of the depth dose distribution (2-25.5 mm) along the transverse axis of a single seed were performed in water equivalent phantom material. The transverse axis attenuation and geometry factor F(r) was obtained by applying a least squares fit to the measured data. Based on the resulting radial dose function, a computer program was developed which calculates dose distributions within the eye for arbitrary loading and placement of the eye plaque. The computational results were verified by TLD measurements in an eye phantom.
Medical subject headings
- Brachytherapy
- Eye Neoplasms
- Iodine Radioisotopes
- Radiotherapy Planning, Computer-Assisted