Nucleic acid assay system for tier II laboratories and moderately complex clinics to detect HIV in low-resource settings.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 20225946.
- Also identified by DOI 10.1086/650388 and PMC identifier 2943734.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There is a clear need for an instrument-free molecular diagnostic system for detecting human immunodeficiency virus type 1 (HIV-1) RNA or DNA that can be used in developing countries. Such a test could be used for early diagnosis of HIV-1 infection during infancy and could serve as a surrogate end point for vaccine trials. We developed the IsoAmp HIV-1 assay (BioHelix Corporation), which targets the HIV-1 gag gene with use of isothermal reverse-transcription helicase-dependent amplification chemistry. The IsoAmp HIV assay uses a disposable amplicon containment device with an embedded vertical-flow DNA detection strip to detect the presence of HIV-1 amplicons. The vertical-flow DNA detection strip has a control line to validate the performance of the device and a test line to detect the analyte. The analyte is detected by a sandwich immunoassay for reporter moieties on a capture probe and a detection probe. The control line consists of the detection probe reporter moiety conjugated to the vertical-flow DNA detection strip. The preliminary limit of detection of the IsoAmp HIV assay was evaluated by testing serial dilutions of HIV-1 armored RNA (Assuragen). We found that 21 (75%) of 28 assays yielded positive results when 50 copies of HIV-1 armored RNA were input into the IsoAmp HIV reaction.
Medical subject headings
- HIV
- HIV Infections
- Molecular Diagnostic Techniques
- Nucleic Acid Amplification Techniques