Pacemaker failure on induction of anaesthesia.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 2025481.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A patient with a permanent pacemaker presented for repair of a strangulated hernia. During induction of anaesthesia, the pacemaker generator stopped discharging, thus causing cardiac arrest. The likely cause of the generator failure was inhibition by suxamethonium-induced muscle fasciculations. Following defibrillation, and increase in stimulation threshold necessitated urgent insertion of a transvenous pacing system. It is suggested that, when suxamethonium is to be used in a patient with a permanent pacemaker, consideration should be given to reprogramming the pacemaker to asynchronous mode before induction of anaesthesia. If a patient with a pacemaker requires defibrillation, an acute increase in stimulation threshold may result and cause loss of capture. Rapid insertion of a transvenous pacing system may be necessary.
Medical subject headings
- Anesthesia, General
- Heart Arrest
- Pacemaker, Artificial
- Succinylcholine