Combined microwave heating and surface cooling of the cornea.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2026437.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated a nonsurgical means of reshaping the cornea to correct hyperopia, keratoconus, or myopia. The object was to heat the central stroma of the cornea to the shrinkage temperature of collagen, 55-58 degrees C. The heating device was an open-ended, coaxial, near-field applicator driven at 2450 MHz; it incorporates cooling of the cornea surface by flow of saline. We investigated the system theoretically by computing the 2-D, axisymmetric temperature distribution with the finite element method. We investigated the system experimentally by heating excised steer corneas. Histology showed the system could shrink the stroma to a depth of 0.6 mm while sparing the epithelium in 75% of cases; the diameter of shrinkage was 1.3 mm. Theory predicted a significantly deeper and narrower region of shrinkage than was observed.
Medical subject headings
- Cornea
- Microwaves